App DB | Sorbent Technologies, Inc.
Home
sorbtech.com
Back
Contact an Account Manager
Separation of PAHs under isocratic conditions
Summary
Type:
HPLC
Application #:
101650
Catalog#:
720923.40 - MN HPLC Column, Nucleosil C18 PAH, 100A, 5um, 4x150mm
Detection:
UV 260 nm
Conditions:
eluent: acetonitrile / water (90 : 10, v/v),
flow rate 1.0 ml/min,
pressure 45 bar,
temperature 27 °C,
injection volume 5 μl.
Authors:
Macherey-Nagel
Source:
LC Department
Details
Legend: naphthalene (18 µg/ml), acenaphthylene (44 µg/ml), acenaphthene (44 µg/ml), fluorene (9 µg/ml), phenanthrene (7 µg/ml), anthracene (11 µg/ml), fluoranthene (11 µg/ml), pyrene (7 µg/ml), benz[a